This model contains resistors, capacitors and inductors which are required for calibrating LCR Meters and can switch connections for measurement frequency, measurement voltage and calibration of DC BIAS voltage manually and automatically. As this compact and light weight unit can be directly connected to LCR Meter it is very convenient to calibrate in the field. Also full automatic LCR Meter calibration is available using software.
SettingRange | Frequency | Accuracy | Temperature coefficient | Long Term Drift | Maximum Rating | |
Resistance | 0.1~30Ω | 1kHz | ±0.1% | ±10ppm/℃ | ±300ppm/year | 0.5W |
30Ω~200kΩ | ±0.01% | ±100ppm/year | ||||
200kΩ~1MΩ | ±0.02% | |||||
Capacitance | 1pF~10pF | 1kHz | ±1% | ±50ppm/℃ | ±100ppm/year | 15Vrms |
10pF~100pF | ±0.5% | |||||
100pF~10μF | ±0.05% | |||||
10μF~100μF | 120Hz | ±0.2% | 3Vrms | |||
1mF | ±0.4% | 180mA | ||||
10mF | ±1% | 500mA | ||||
100mF | 1A | |||||
Dissipation Factor | 0.001~1.000(1000pF~10μF) | 50Hz~1kHz | Less than±0.0001 | ±200ppm/℃ | ±300ppm/year | 15Vrms |
0.001~1.000(10μF~1mF) | 50Hz~120Hz | 3Vrms | ||||
Inductance | 10μH~10mH | 1kHz | ±0.5% | ±50ppm/℃ | ±100ppm/year | 10Vrms |
10mH~1H | ±250ppm/℃ | ±500ppm/year | 3Vrms |